Specifications
34 CHAPTER 4: CONFIGURATION AND SETUP
087-0048 REV B SCOTT SAFETY
Connecting the IR
Connect to the Test
Station
This section describes connecting the IR Connect to the Test Station to use the Test
Station Software.
The purpose of this hardware configuration is to configure multiple monitors at a
time.
Required items:
This assumes you have already installed the Test Station Software.
• The IR Connect and its USB Cable
• A PC
• Test Station with AC Power Adapter
1 Plug the USB Cable into the PC’s USB port and the other end into the IR Connect.
2 Place the IR Connect in the left most slot labeled “Unit 1” of the Test Station.
Ensure the two IR ports are facing each other and square. See Figure 26.
The IR Connect only operates in the “Unit 1” slot.
Figure 26 Connecting the IR Connect to a PC & the Test Station
3 To power on the Test Station, press the Bump Test Button.
4 Ensure the Test Station powers up properly. The Power LED lites Green - Red and
then Green. The Unit1 to Unit4 LEDs lites Orange - Red and then Off. If the AC
Adapter is used and battery is charging, then the 4 Unit LEDs lite in sequence from
left to right.
The Test Station is equipped with an automatic power shut off after sitting idle for 120
seconds. Power LED lites Green while On. This parameter is configurable.
5 Open the Test Station Software that was already installed.
6 After proper IR Connect placement, ensure IR is selected and click READ. A
successful connection results in a green bar in the lower right hand corner and the
Unit 1 LED flashes Green. See Figure 27.
IR Connect
Unit 1