Specifications
Configuring the IR Connect Software Parameters 29
SCOTT SAFETY 087-0048 REV B
Changed parameters appear in Green prior to WRITE DEVICE execution.
Table 12 IR Connect Software Parameters
ITEM DETAILS
High Alarm/Low
Alarm Set Points
Allows adjusting the low/high alarm set point from the factory defaults.
• For H
2
S:
Default low/high =10/15PPM
Range =1 to 100 PPM
• For CO:
Default low/high =35/200PPM
Range =1 to 300PPM
• For O
2
:
Default low/high=19.5/23.5% Volume
Range=1 to 30% Volume
Note: Ensure the set alarm limits do not exceed the local or site
guidelines. Some sites may have different set points than others.
Calibration Interval Displays the following values based on the monitor’s type:
• For H
2
S: Default =0days
• For CO: Default =0days
• For O
2
: Default =30days
Note: The O
2
monitor factory default Calibration Interval is set to
30days. Thus, every 30 days, the monitor displays a flashing CAL
screen, indicating the monitor needs Calibrating.
Bump Interval Setting a value automatically displays “BUP” on the monitor’s LCD
when the monitor has not been Bump Tested within the time frame and
remains until cleared. This option aids in bump test accountability.
Default=0 days
Range=0 to 365 days
For details on clearing this parameter, See “Clearing a Bump Test
Interval Alarm Alert” on page 61.
User ID Allows entering up to six (6) characters to identify each individual
device. Displays in two screens when all 6 characters are used.
Note: Tab down to this field to select prior to entering ID.Otherwise
field will not accept entry. Also, some alphabetic characters do not
display well on the segmented display.
Self-Test Interval Allows adjustments to this item.
Default=20hours
Range=6 to 24hours
Display Sets detector display; Sensor/Life Remaining, Sensor Reading, or Life
Remaining. Sensor/Life Remaining mode shows life remaining instead
of zero. When it’s not reading zero, then it shows the gas reading.
Default= Sensor/Life Remaining
Hide Bump Due
LED Flashing
When checked, disables the monitor’s LEDs from flashing when due
for a Bump Test.
When unchecked, the monitor’s LEDs flash when due for a Bump Test.
Default= Unchecked