Specifications
Specifications 81
SCOTT SAFETY 087-0048 REV B
Table 27 IR Connect Specifications
IR CONNECT SPECIFICATIONS
Range 2.0 to 16.0” (50 to 406mm)
Power USB (5V)
Operating System Compatible with MS Windows XP, Vista and Windows 7
Reporting Event logs from the monitor and Firmware upgrades to the
monitor
Readings &
Configurations
Alarm set points to the monitor
Dimensions 3.8Hx2.5Wx1.1”D (95Hx63Wx28mmD)
Weight 2.7oz (76g)
Operating
Temperature
50 to 104°F (10 to 40°C)
Operating Humidity 5% to 95% RH, Non-Condensing
Table 28 Test Station Specifications
TEST STATION (PORTABLE & STATIONARY) SPECIFICATIONS
Test Performed Bump Test & Calibration
Test Capacity Up to four (4) monitors simultaneously
DC Battery 12VDC Lithium Ion rechargeable
AC Power Adaptor • Equipped with adapter
• Input Voltage: 100 to 140VAC, 47 to 63Hz
• Input Current: 450mA @ 100VAC
Test Station DC Input Power:12VDC, 1.5A Max.
Bump Time <30Seconds
Calibration Time <2Minutes
Regulator Flow Rate
(Portable only)
0.3LPM
Bump Cycles/DC
Battery Charge
1,000 (~33hrs)
Memory Ships with a 2GB USB Standard Pen Drive Installed (Larger
memory size may be used)
LEDs Power, Unit1, Unit2, Unit3 and Unit4
Buttons Bump Test and Calibration
Information Stored Bump/Calibration Logs, Individual Monitor Event Logs,
Firmware, and Unit Configurations
Configuration Unit settings (configuration and Firmware) may be
programmed/upgraded via the Test Station