User's Manual
PROOF DIAGNOSTICS COVID-19 TEST SYSTEM
26 27
Technical Specications
PROOF LAB SPECIFICATIONS
Dimensions 5.5 width x 5.5 length x 6.7 height (140 x
140 x 170 mm)
Weight < 2.5 lbs (< 1.13 kg)
Power Supply Specications Low voltage power supply +12V,
24Watts
Operating Temperature 64 to 86°F (18 to 30°C)
Operating Humidity 15% to 80% relative humidity
Operating Altitude Less than 2000m
COVID-19 TEST KIT SPECIFICATIONS
Shelf Life TBD
Storage Temperature 64 to 86°F (18 to 30°C)
Error Alerts
ERROR STATE 1 – POWER ON ERROR
What to do
1. Open Proof Lab lid.
2. If Cartridge is present, remove
Cartridge. Note: If Cartridge is
punched, record test result and
discard. If not punched, record
“no result” and discard.
3. Close lid. Wait 30 seconds for
Proof Lab to return to Ready
state. If the display returns to
the Ready state, start a new
test.
4. If error remains, contact
Technical Support at (tbd).
ERROR STATE 2 – INVALID
What to do
1. Record Invalid result, then
open lid.
2. Remove Cartridge from the
Proof Lab and note if the result
label is punched as Invalid. If
so, record test result.
3. Close lid. Wait 30 seconds
for Proof Lab to return to
Ready state. If the display
returns to the Ready state,
start a new test.
4. If error remains, contact
Technical Support at (tbd).