User's Manual
PROOF DIAGNOSTICS COVID-19 TEST SYSTEM
14 15
19. Firmly close the lid of the Proof Lab.
When the lid is closed, the Proof Lab will
detect the presence of the Cartridge.
20. When the Cartridge is detected, the test
will begin. DO NOT OPEN THE
LID WHILE THE TEST IS RUNNING� This
will invalidate the test. If the Cartridge is
not detected, Proof Lab will return to the
Ready state as indicated by the white
LED ring shown in ”Prepare Proof Lab”
section on page 8. Start again by rmly
re-inserting the Cartridge into the
Proof Lab.
Test Result Interpretation
When the test is complete, the result will be shown on both the Proof Lab
display and the Test Cartridge.
Proof Lab Display
When the test is complete, the result will appear on the Proof Lab display.
Record the test result shown on the Proof Lab.
Test Cartridge
The test result is also recorded on the Test Cartridge
Negative Positive Invalid
Powering Off the Reader
To power off the Proof Lab: Check to make sure that no Test Cartridges have been
left in the Proof Lab. Unplug the power cord from the main power outlet.
21. When the test is complete, the result will appear on the Proof Lab display.
Record the test result shown on the Proof Lab. (For further information, see
“Test Result Interpretation” on page 15.)
22. Open the lid of the Proof Lab and remove the Cartridge.
23. The test result will also be punched on the Cartridge. (For further
information, see “Test Result Interpretation” on page 15 and “Reporting
Test Results” on page 19.)
24. Record the result and discard the Cartridge in a biohazard waste bin.
Invalid
Negative
Positive