Service manual

Introduction Precautions
vi 0070-10-0634 Panorama® Service Manual
WARNING: Do not attempt to load any devices or device drivers onto
Panorama
®
Central Station. If the user connects or attempts
to connect any equipment Panorama
®
Central Monitoring
System may not operate as intended.
WARNING: Only qualified and trained service personnel or Service
personnel should attempt to service equipment. Service is
defined as any activity requiring the cover to be removed
for internal adjustments, parts replacements, repairs or
software upgrades of any kind to insure compatibility.
WARNING: To insure compatibility with the operating system and
applications software, use only the supplied and/or
approved components to repair any part of the Panorama
Patient Monitoring Network. Use of unauthorized software,
devices, accessories, or cables other than those approved
may render the application unsuitable for medical patient
monitoring. It may also result in increased electromagnetic
Emissions or decreased Immunity of the system.
WARNING: Do not block or turn down the volume from the maximum
position on the speakers provided with Panorama
®
Central
Monitoring System. Set the volume levels so that alarms can
be heard at all times, as described in the Operation Manual.
WARNING: Be careful not to turn off patient alarms. Turning off patient
alarms can jeopardize patient safety.
WARNING: Do not incinerate batteries, possible explosion may occur.
WARNING: The 18.1" flat panel may tip over, if the display head is
inclined to an angle greater than 45°, backward tilt. If the
user elects to have the display head inclined to an angle
greater than 45°, backward tilt, the flat panel must be
attached to a secure mounting surface via three screw
locations on the bottom of the base.
WARNING: The Panorama Central Station and the Panorama Telemetry
Server must utilize the hospital emergency power system.
Failure to do so will result in loss of monitoring during
extended periods of power failure. The back-up power time
period, for the Panorama Patient Monitoring Network, is
limited.
WARNING: Do not put MPSO (Multiple Portable Socket Outlets, i.e.,
multiple outlet extension cords) used with the Panorama
Central Station System on the floor. Connect only Panorama
Central Station accessories and components to the same
MPSO as the Panorama Central Station. Do not overload
MPSOs. Use only MPSOs that comply with the requirements
of IEC 60601-1-1.
WARNING: Do not clean the monitor while it is on and/or plugged in.
Precautions
CAUTION: This system is intended for use in a hospital or clinical
setting and to be operated by trained and authorized
personnel who are acting on the orders of a physician. Its
purpose is the real time monitoring of a patient's
physiological parameters over an extended time frame.