Service manual
Introduction Connections and Configuration
6 - 6 0070-10-0634 Panorama® Service Manual
For Hardwired Ethernet Connection
Each bedside monitor must have a different IP address: 7.6.6.2, 7.6.6.3, etc.
NOTE: The first 3 digits X.X.X. must match the IP address of the ELAN
address of the Central Station.
1. Enter the Installation Menu by pressing and holding the Discharge key during power
up. Release the Discharge key when the Installation Menu appears.
2. Go to the System Information Sub-Menu.
3. Set Enable Network to Wired.
4. Enter an IP Address that matches the Patient Network IP Address scheme (e.g., 7.6.6.2).
Do not duplicate an existing assigned IP address.
5. Verify that the Subnet Mask ID is 255.255.255.0.
6. Check the Device ID number. This number will be entered into the Central Station's
equipment list.
7. Make sure neither serial port is set to:
• VISA with Admit, or
•PatientNet
8. Save the information by pressing Save Current.
9. Turn off the monitor.
For 2.4 GHz Wireless Telemetry Connection
Each bedside monitor must have a different wireless IP address: 7.8.8.1, 7.8.8.2, etc.
NOTE: The first 3 digits (X.X.X.) must match the IP address of the
WELAN address of the Telemetry Server.
1. Enter the Installation Menu by pressing and holding the Discharge key during power
up. Release the Discharge key when the Installation Menu appears.
2. Go to the System Information Sub-Menu.
3. Set Enable Network to Wireless.
4. Enter a Wireless IP Address that matches the Telemetry Server address scheme (i.e.,
7.8.8.2). Do not duplicate an existing assigned Wireless IP address.
5. Verify that the Wireless Subnet Mask ID is 255.255.255.0.
6. Enter the Network Name assigned to the Wireless Network.
7. Check the Device ID number. This number will be entered into the Central Station's
equipment list (software versions below 8.XX only).
8. Make sure neither serial port is set to:
• VISA with Admit, or
•PatientNet
9. Save the information by pressing Save Current.
10. Turn off the monitor.