Panorama® Service Manual 0070-10-0634 5 - 7
Preventative Maintenance System Maintenance
Lead Wire Set and Lead LED Testing.
1. Press the Check Button.
2. Verify illumination of all connected Lead LEDs.
If one or more LEDs are not illuminated, replace the lead wire set and repeat the test.
If one or more LEDs are still not illuminated, replace the Telepack.
Lead Select Testing (3 Wire Lead Set Only)
1. Press the Lead Select button:
Verify the Lead LED pattern changes every time the Lead Select button is pressed.
If the Lead Pattern LEDs do not change, replace the Telepack.
NOTE: This feature does not function with no lead wire set attached
or with a 5-wire lead set.
Additional Button Testing
NOTE: The following buttons can only be tested if the Telepack has
been admitted to the system.
1. Nurse Call Button (if Nurse Call is active at the Central Station).
Press the Nurse Call button and verify the Nurse Call message is displayed on the Patient
tile for that channel.
2. Attendant Present Buttons (if Nurse Call is active at the Central Station).
Press the two Attendant Present buttons and verify the nurse call message is cleared from
the Patient Tile for that channel.
Record Button
Press the Record Button and verify a printout is completed at the Central Station for that
channel.