Panorama® Service Manual 0070-10-0634 5 - 5
Preventative Maintenance System Maintenance
5.5 System Maintenance
5.5.1 Preventative Maintenance
The Preventative Maintenance procedure for the Panorama Patient Monitoring Network will
require periodic cleaning, inspection and testing. Use the following tables for each
maintenance procedure:
Display Monitors
ACTIVITY
AREA ACTION
TECHNICAL
LEVEL METHOD SCHEDULE
Screen Clean User See Operators
manual Section
11.2
As required
Touch
Screen
Calibrate User Built-in system
software
As required
Case Clean Bio-Medical/
Clinical Engineer
Cloth dampened
with mild
detergent*
Semi-Annually
Cables Inspect Bio-Medical/
Clinical Engineer
-- Semi-Annually
Central Station and Telemetry Server
ACTIVITY
AREA ACTION TECHNICAL LEVEL METHOD SCHEDULE
Cooling Fans Clean Bio-Medical/
Clinical Engineer
Forced air Semi-Annually
Power Supply
Fans
Clean Bio-Medical/
Clinical Engineer
Forced air Semi-Annually
Floppy Drive Clean Bio-Medical/
Clinical Engineer
Forced air Semi-Annually
Connections Inspect Bio-Medical/
Clinical Engineer
-- Semi-Annually
Case Inspect for
damage
Bio-Medical/
Clinical Engineer
-- Semi-Annually
Laser Printers
ACTIVITY AREA ACTION TECHNICAL LEVEL METHOD SCHEDULE
Connections Inspect Bio-Medical/
Clinical Engineer
-- Semi-Annually
Case Clean Bio-Medical/
Clinical Engineer
Cloth
dampened with
water *
As required
Printing areas Clean Bio-Medical/
Clinical Engineer
Lint-free cloth When toner
cartridge is
changed