Verify the "1" at the bedside monitor. This indicates communication with the Access
Point.
FIGURE 2-3 Bedside Monitor (Wireless)
The Status LED on the Access Point is blinking on/off. If the LED is steady on, steady off
or flashing irregularly, replace the Access Point.
Verify the Wired and Wireless LAN LEDs on the Access Points. Verify the link LEDs are
on steady. Activity LEDs are flashing when there is traffic.
Verify the Power LED is illuminated on the Power Distribution power supply. Verify 15
VDC at the output of the power distribution system.
Verify power at the POE Modules at the Access Point and at the Panorama Wireless
Telemetry Server locations. Verify the green LED is illuminated. This indicates 15 VDC
power to the unit.
Verify the connection of the CAT5 cable between the POE Module and the Patch
Panel.
Verify the connections between the Patch Panel, the POE Module, the switch, and the
Power Distribution/Power Supply. Verify AC and DC power at these units and verify
the Status LEDs.
Verify the correct connections at the Panorama Wireless Telemetry Server. The Patient
Network connection is not at the same location on the Panorama Wireless Telemetry
Server as it is on the Central Station.