Owner's manual
6
DC VOLTAGE
Overload Protection: 250V rms. For 200mV range and 600V dc
or rms. ac for other ranges.
DC CURRENT
Overload Protection: F 200mA/250V fuse. (10A range unfused)
*303-150NCS DC 200µA Excepted
AC VOLTAGE
Overload Protection: 600V dc or rms. ac for all ranges.
Frequency range: 40Hz to 400Hz. Response: Average respond-
ing, calibrated in rms. of a sine wave.
DIODE & CONTINUITY(*903-150NAS Excepted)
Range Resolution Accuracy
200mV 100µV ±0.5% of rdg ± 2 digits
2V 1mV ±0.5% of rdg ± 2 digits
20V 10mV ±0.5% of rdg ± 2 digits
200V 100mV ±0.5% of rdg ± 2 digits
600V 1V ±0.8% of rdg ± 2 digits
Range Resolution Accuracy
200µA 0.1µA ±1% of rdg ± 2 digits
2mA 1µA ±1% of rdg ± 2 digits
20mA 10µA ±1% of rdg ± 2 digits
200mA 100µA ±1.5% of rdg ± 2 digits
10A 10mA ±3% of rdg ± 2 digits
Range Resolution Accuracy
200V 100mV ±1.2% of rdg ± 10 digits
2600V 1V ±1.2% of rdg ± 10 digits
Range Description
If continuity exists (about less than 1.5kΩ), built-in buzzer
will sound.
Show the approx. forward voltage drop of the diode.