Operation Manual

89
© Prognosys Medical Systems Private Limited│ X-Ray Products
Radiated RF
IEC 61000-4-3
in ISM bands
between
0,15 MHz and 80
MHz
80 % AM at 1
kHz
3 V/m
80 MHz 2,7
GHz
80 % AM at 1
kHz
in ISM bands
between
0,15 MHz and 80
MHz
80 % AM at 1
kHz e)
3 V/m
80 MHz 2,7
GHz
80 % AM at 1
kHz
HC including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1,2
P
d = 1,2
P 80MHz to 800 MHz
d = 2,3
P 800MHz to 2,7 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.