HC be powered from an
uninterruptible power supply
or a battery.
Power
frequency(50, 60
Hz) magnetic field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz
The PRORAD V14 CLARITY,
PRORAD V14 HC, PRORAD
V17 CLARITY, PRORAD V17
HC power frequency
magnetic fields should be at
levels characteristic of a
typical location in a typical
professional healthcare
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Manufacturer’s declaration-electromagnetic immunity
The PRORAD V14 CLARITY, PRORAD V14 HC, PRORAD V17 CLARITY, PRORAD V17 HC is intended for use
in the electromagnetic environment (for professional healthcare) specified below.
The customer or the user of the PRORAD V14 CLARITY, PRORAD V14 HC, PRORAD V17 CLARITY,
PRORAD V17 HC should assure that it is used in such and environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment-guidance (for
professional healthcare environment)
Conducted RF
IEC 61000-4-6
3 Vrms:
0,15 MHz – 80
MHz
6 Vrms:
3 Vrms:
0,15 MHz – 80
MHz
6 Vrms:
Portable and mobile RF communications
equipment should be used no closer to any
part of the PRORAD V14 CLARITY, PRORAD
V14 HC, PRORAD V17 CLARITY, PRORAD V17