Operation Manual
87
© Prognosys Medical Systems Private Limited│ X-Ray Products
synthetic material, the
relative humidity should be
at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
+ 2kV for power supply
lines
+ 1kV for input/output
lines
+ 2kV for power supply
lines
+ 1kV for input/output
lines
Mains power quality should
be that of a typical
professional healthcare
environment.
Surge IEC
61000-4-5
+ 0.5kV, +1kV line(s) to
line(s)
+ 0.5kV, +1kV,+ 2kV
line(s) to earth
+ 0.5kV, +1kV line(s) to
line(s)
+ 0.5kV, +1kV,+ 2kV
line(s) to earth
Mains power quality should
be that of a typical
professional healthcare
environment.
Voltage Dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
Voltage dips:
0 % U
T
; 0,5 cycle
0 % U
T
; 1 cycle
70 % U
T
; 25/30 cycles
Voltage interruptions:
0 % U
T
; 250/300 cycle
Voltage dips:
0 % U
T
; 0,5 cycle
0 % U
T
; 1 cycle
70 % U
T
; 25 cycles
Voltage interruptions:
0 % U
T
; 250 cycle
Mains power quality should
be that of a typical
professional healthcare
environment. If the user of
the PRORAD V14 CLARITY,
PRORAD V14 HC, PRORAD
V17 CLARITY, PRORAD V17
HC requires continued
operation during power
mains interruptions, it is
recommended that the
PRORAD V14 CLARITY,
PRORAD V14 HC, PRORAD
V17 CLARITY, PRORAD V17