Operation Manual

86
© Prognosys Medical Systems Private Limited│ X-Ray Products
Manufacturers declaration-electromagnetic immunity
The PRORAD V14 CLARITY, PRORAD V14 HC, PRORAD V17 CLARITY, PRORAD V17 HC is intended for use
in the electromagnetic environment (for professional healthcare) specified below.
The customer or the user of the PRORAD V14 CLARITY, PRORAD V14 HC, PRORAD V17 CLARITY,
PRORAD V17 HC should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment-guidance
(for professional healthcare
environment)
Electrostatic
discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
RF emissions CISPR 11
Group 1
The PRORAD V14 CLARITY, PRORAD V14 HC, PRORAD
V17 CLARITY, PRORAD V17 HC uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11
Class A
The PRORAD V14 CLARITY, PRORAD V14 HC, PRORAD
V17 CLARITY, PRORAD V17 HC is suitable for use
in all establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations
/flicker emissions IEC
61000-3-3
Compliance