Operation Manual

80
© Prognosys Medical Systems Private Limited│ X-Ray Products
MTF @ 2 lp / mm
0.17
0.23
MTF @ 3 lp / mm
0.07
0.11
DQE @ 0.5 lp / mm
0.23
0.31
DQE @ 1 lp / mm
0.22
0.27
DQE @ 2 lp / mm
0.12
0.16
DQE @ 3 lp / mm
0.06
0.09
*The above specifications have been verified by Prognosys; The physical characterization of the
systems was obtained with the standard beam conditions RQA5 (in accordance with IEC
61267:1994).
A.1.7 Radio Waves
Wireless specifications for the flat panel detector and the access point are as follows.
Item
Specification
Wireless specification
IEEE802.11 ac / a/g/n
Transmit frequency
5.2, 5.3, 5.6, 2.4 GHz
Antenna
2
Date rate
866Mbps (802.11ac)
Authentication
WPA/WPA2-PSK
AP mode
Work only in 2.4Ghz
- Note: Use of 2.4GHz: ch11 for N.America, ch13 for Europe (ETSI)
- Transmit frequencies available vary, depending on the country.
- Radio waves available outdoors vary, depending on the country where the system is used.
- When the PRORAD and any other wireless equipment are operating on the same frequency
channel in a hospital, it may take time to show an image on the image processing unit monitor.
- This equipment uses wireless LAN (WLAN) radios for transferring images. The WLAN power