Operation Manual

77
© Prognosys Medical Systems Private Limited│ X-Ray Products
A.1.5.3 Performance Operating
Performance is guaranteed in the operating range below.
Item
Min.
Typ.
Max.
Temperature
15°C
25°C
30°C
Humidity
15%
80%
Pressure
700mBar
1013mBar
1060mBar
Distance to calibration
temperature
-12°C
0°C
+12°C
Weight (uniform load)
300 kg
Weight (local load on 4 cm
diameter disc)
120 kg
Vibrations (10 -200 Hz / 3 axis )
2 g
Drop (The PRORAD-V14
CLARITY/G, PRORAD-V17
CLARITY/G remains at full
performance after a drop test of
following height)
70 cm
_________________________________________________________________________________
Caution
When the flat panel detector is used in high temperature condition for long period of time, it may
cause image artifacts and/or failure of the device.
_________________________________________________________________________________
A.1.6 Image Performance