Operation Manual

76
© Prognosys Medical Systems Private Limited│ X-Ray Products
Pressure
2
700mBar
1060mBar
Battery storage temperature
3
-10°C
50°C
- note1: Non-condensing
- note2: 700mBar is equivalant to 3000m in altitude
- note3: Less than 1 month at 50°C
A.1.5.2 Functional Operating
Within the range of functional operating conditions, the PRORAD V14 CLARITY, PRORAD V14 HC,
PRORAD V17 CLARITY, PRORAD V17 HC are safe and functional.
Functional means that the detector is able to deliver an image, whereas the image quality and
electro-optical performances are not guaranteed. It also means that, except if mentioned explicitly,
any perturbation occurring within those conditions are reversible and not existing anymore when
the detector is turned back within the performance operating conditions.
Item
Min
Typ.
Max
Temperature
5°C
25°C
35°C
Humidity
15%
80%
Pressure
700mBar
1013mBar
1060mBar
Weight (uniform load)
300 kg
Weight (local load on 4 cm
diameter)
120 kg
Vibrations (10 -200 Hz / 3 axis )
2 g
Drop height (the reversibility
to full performance is not
guaranteed for this item)
70 cm