4.5.3 Standard Configuration
A. Other Essential System Components necessary to perform Digital Radiography
PRORAD (Supporting SDK Interface) can be used in combination with the image processing unit
provided by System Integrator.
The X-ray equipment is composed of X-ray high voltage generator and X-ray console, etc. The
hand switch is connected with the console for the exposure. At first, when making an exposure,
press the Prep switch only of the hand switch. After confirming that the X-ray equipment is
ready System Integrator of X-rays, press the Exp. switch of the hand switch.
The Access Point is required system, when PRORAD is used in wireless communication mode.
B. A standard configuration of this device is as below.
PRORAD (Supporting SDK Interface):
Flat panel sensor : PRORAD V14 CLARITY / V14 HC / V17 CLARITY / V17 HC.
The control software is installed and used in a Windows PC (the image processing unit) prepared by
the System Integrator.
This variation provides images, which diagnostic image process is not implemented, to the image
processing unit.
An X-ray automatic detection function is used for PRORAD V14 CLARITY / V14 HC / V17 CLARITY /
V17 HC. This function could detect a start timing of an X-ray exposure using readout signals in the