4.5 Correction
_____________________________________________________________
Cautions
This procedure should only be done by suitably-trained service engineer.
_____________________________________________________________
Cautions
If the exposure for the corrected images are collected by wrong exposure setting, for example, (A).
Over-exposure than or under-exposure (B). Foreign object is put on the surface (C) collimator of
Xray tube is not open to the full field of FPD. It may lead to image artifact.
_____________________________________________________________
Cautions
Do not expose to any X-ray during dark image collection.
_____________________________________________________________
4.5.1 Collect images
When the program is executed, the corrected images are collected in sequence. The images with
X-ray radiation and without X-ray radiation are collected in reasonable amounts according to the
environment. At least 5 flat field images are recommended for calibration. Radiation quality of the
light image shall be in the range of 30000±3000lsb.(The exposure condition is adjustable based on
each customers.) The images with X-ray radiation and without X-ray radiation are collected in
reasonable amounts according to the environment. Next, the offset map (D) and the gain map (G)
are calculated. And use the following algorithm to correct for raw images.
- C = corrected image
- R = raw image
- F = flat field image
- D = dark field or dark frame
- m = image-averaged value of (F-D)
- G = Gain =