Operation Manual
58
© Prognosys Medical Systems Private Limited│ X-Ray Products
Width: The width of image
Height: The height of image
Connection Type: Wireless or Wired
Net Speed: The strength of signal
Diagnostic unit (The group is collapsed by default, click + to extend)
Battery: Main Battery information
Secondary Battery: Second Battery information
G-Sensor
Humidity
Pressure
Motion-Sensor
Angular-Sensor
FPD information
Mode: The operation mode
State: Current state of operation
Source: Power source
Subsystem: Subsystem status
Alarm flags:
Image module: Alarm caused by image module
Storage: Alarm caused by storage failure
Ethernet: Alarm caused by Ethernet interface failure
Wireless: Alarm caused by Wireless interface failure
Self-diagnosis: Alarm caused by software self-diagnosis failure
Control Host: Host name of the PC currently controls the operation of FPD
Error (The group is collapsed by default, click + to expand)
Error Code: A 32-bit mask of following error flags (0x00000000 means no error)
ROIC 0~17: Failure on checking ROIC index N (0~17) configuration (bit 0 ~17)
Gate: Failure on checking gate (bit 24)
ROIC init: ROIC initialization timed out (bit 25)
AED: Failure caused by AED operation (bit 26)
DMA: DMA timed out (bit 27)
Power: Power system failure (bit 28)
Image module: Image module timed out (bit 29)
Panel: Panel configuration error (bit 30)
Driver: Software driver error (bit 31)