Operation Manual
57
© Prognosys Medical Systems Private Limited│ X-Ray Products
None: No offset image OAX: Offset image after exposure
OBX: Offset image before exposure
Clear Before OBX (Clear1): Discharge before acquiring OBX.
Delay between Clear1 and OBX: If this option is checked, there will be an offset window delay
between Clear1 and OBX. Otherwise the OBX is acquired immediately after Clear1.
Clear Before Exposure (Clear2): Discharge before X-Ray exposure.
X-Ray Window (ms): X-Ray window in milliseconds. [2.5ms~3s]
Offset Window (ms): Offset window in milliseconds. [2.5ms~3s]
Caption Bar
Display the latest message of output panel.
Discovered FPDs Panel
This panel displays a list of FPDs that can be discovered in the local area network.
Every list item has two subitems
Ethernet IP of FPD
Wi-Fi IP of FPD
FPD Information / Configuration
This Panel contains tabs of FPD information and configurations
Properties
The properties page contains following information of FPD
FPD specification
Serial: Serial number of FPD
Model: Model name of FPD
System version: Firmware version of main system
Subsystem version: Firmware version of subsystem
FPGA version: Bitstream version of FPGA