Operation Manual

56
© Prognosys Medical Systems Private Limited│ X-Ray Products
Mode
Start: Make FPD operate in selected exposure mode.
Stop: Make FPD stop selected exposure mode and return to Idle in operation state.
Reset: Reset the image module (Clear some alert flags).
Mode: Choose exposure mode (AED / AED repeat / Soft Sync / Acquisition / Discharge).
AED: Automatic exposure detection is activated with Utility, and to detect the X-ray and capture
image directly.
AED repeat: Automatic exposure detection is activated with Utility at first time, the following
activation will be done with panel itself and saved in the storage of FPD. *
Soft Sync: Synchronizing detector and radiation system to exposure.
Acquisition: Capturing image without X-ray radiation.
Discharge: Reset the panel.
* Alarm: Please unchecking “don’t save to eMMC” while using AED repeat mode, or the images
would not be saved properly once the temporary storage of DDR is full.
User Key: A number for user key generation. The number increases by 1 after each exposure.
Download Raw: Download the raw image to the target folder.
Download Offset: Download the offset image to the target folder.
Download Corrected Image: Download the corrected image to the folder named “Corrected”. (It
will automatically create the folder named corrected but the calibration is depending on the
availability of the maps or offset configuration.)
Preview Image: Activate the function of preview image and download the preview image to the
folder named “ThumbNail”. (The saving priority is the first if checked)
Don’t Save to eMMC:
When this option is not checked, the image is saved in the non-volatile storage of FPD.
When this option is checked, the image is only stored in buffer of FPD and will be overwritten by
subsequent exposure image.
Acq Repeat: The numbers of images would automatically be collected. (Only for images without
exposure.)
Bulk Size: Determine the data size to be divided to transfer.
Update Maps: Renew the maps in the CalibrationMap folder to process the calibration
Sub-folder: Provide the options to download the images in the corresponding subfolder.
Override Naming Rules: Name the images as the fields, File Name and Starting Index, shown in
Image List section.
Exposure Configuration (Please refer to AED / Acquisition section for more detail)
The settings cached when user presses Start button. If user changes the settings, the altered
settings is applied the next time user presses Start (after Stop the previous operation)
Offset: