Operation Manual
29
© Prognosys Medical Systems Private Limited│ X-Ray Products
Although the flat panel detector conforms to IPX6, no warranty is given as to the prevention of
water intrusion in the flat panel detector. If the flat panel detector is splashed with water, wipe off
moisture and ensure that the flat panel detector is completely dry before use.
As the cables of the equipment are long, be careful not to entangle the cables during use. Also, be
careful not to trip over the cables. Falls could result in injury.
Follow the specified procedure when turning off the equipment. Otherwise, the flat panel detector
could be damaged by thermal shock.
External equipment intended for connection to signal inputs, signal outputs or other connectors
shall comply with the relevant product standard e.g. IEC 60950-1 or IEC 62368-1 for IT-equipment
and the IEC 60601-series for Medical Electrical Equipment. In addition, all such combinations –
Medical Electrical Systems – shall comply with the safety requirements stated in the general
standard IEC 60601-1, edition 3, clause 16. Any equipment not complying with the leakage current
requirements in IEC 60601-1 shall be kept outside the patient environment i.e. at least 1.5 m from
the patient support or shall be supplied via a separation transformer to reduce the leakage
currents.
Any person who connects external equipment to signal inputs, signal outputs or other connectors
has formed a Medical Electrical System and is therefore responsible for the system to comply with
the requirements. If in doubt, contact qualified medical technician or your local representative.
Do not place FPD near strong EMI sources, it might cause line defect in the image, and patient need
to retake x-ray.
Careful when holdinig the FPD, drop or impact will result in water come inside causing short circuit,
which might have patient retake x-ray.
Do not operate the FPD beyond the limited temperature, patient might need to retake or harmed