Operation Manual
27
© Prognosys Medical Systems Private Limited│ X-Ray Products
Regarding the storage of the battery, first of all, the best storage capacity of the battery is about
40~50%. Before the installation, when the battery and the product are still in the warehouse, it is
recommended that the main battery and the internal battery must be checked every 3 months to
charge.
The main battery needs to be charged to 50%, two bars of the charging stand, and it takes about
2 hours to charge from 0%. ; The internal battery is also charged to 50%, you can use the UI to
check the level of the internal battery, it takes about 1 hour to charge from 0%.
After charging, make sure that the hotswap function operates normally, please unplug the main
battery and press the power button to test the internal battery power at least 3 lights, and the
internal battery power should be more than 50%
_________________________________________________________________________________
2.14 Warnings for Pediatric Use
_________________________________________________________________________________
Warning
If the exposure conditions for average excessive radiation exposure.
Studies show that children are more radiosens of developing cancer compared to adults exposed to
the same dose of ionizing
Accordingly, in pediatric use, special attention needs to be paid to avoid
exposure.
Based on the clinical application, pathological conditions of the patient, patient size, and
anatomical imaging region, adjust the exposure conditions to use the minimum amount of