- Where the equipment is subject to frequent or excessive vibration/shock.
- Where the equipment is exposed to direct sunlight.
Use the equipment on a flat place. If the equipment falls, it may cause damage to the equipment or
personal injury.
When you move the equipment, place it in the cassette storage box of a mobile X-ray unit or hold it
by hand to prevent it from falling. If the cart is used to move the equipment, place it horizontally.
For veterinary or mobile applications, contact our official dealer or Prognosys Representative.
When the devices are used outdoors in wireless communication mode, contact our official dealer or
Prognosys Representative.
Do not place any object in a place where removal of the power cable is prevented.
To ensure optimal image quality, it is recommended that you do not use the flat panel detector
near devices (motor, transformer, switching supply, etc.) that generate electromagnetic noise.
To ensure optimal image quality, it is recommended that you do not place the cables (power cable,
back-up cable, etc.) of the equipment near devices (motor, transformer, switching supply, etc.) that
generate electromagnetic noise and their cables.
Please use the engineering version of UI to reset the Real-time clock (RTC) every time you
use it, or make sure that the RTC of the FPD is correct when connecting to the radiology
software, to avoid wrong time records in the log.
If you use the engineering version of the UI for single-dose calibration, it is recommended to
use 70kv and the raw image to reach the conditions of 30000±3000lsb to achieve the best
calibration map.
Please use the INCX service tool to update the FPD firmware to avoid update errors.