Configure New Setpoint Mode
New Setpoint Mode
To configure the New Setpoint Mode, press the
checkbox to activate setting, as shown by the
X.
Grey No setpoint change.
Green Active; Change setpoint to
New Setpoint.
Configure New Setpoint
New Setpoint
To configure a New Setpoint when limit is
reached, press the percentage value, as shown
by the X.
Red Border when value reaches limit, the new
setpoint is locked until
Totalizer Reset is
executed.
New Setpoint is the percent value of the FULL
Instrument range.
Examples:
New Setpoint at 0%:
Full Instrument Range: 500 grams/hr
Setpoint: 400 grams/hr
When limit is reached,
New setpoint value: 0 grams/hr
New Setpoint at 50%:
Full Instrument Range: 500 grams/hr
Setpoint: 400 grams/hr
When limit is reached,
New setpoint value: 250 grams/hr
IMPORTANT:
Instrument will not stop flowing after limit is
reached unless
New Setpoint Mode is
active/green AND
New Setpoint is 0%.