Operating instructions
10 © 2013 Proceq SA
6
Continuous/Burst Transmission:
Continues transmitting until the stop icon is pressed.
Records a measurement as soon as a stable signal is
detected.
7
Settings: Enter the settings menu.
8
Stop/Save (Right Button on Transducer):
Stop the current measurement.
Save the current measurement.
Save the current series and continue the measurement.
9
Start/Snapshot (Left Button on Transducer):
Begin the measurement.
Save the current measurement as displayed on the
screen and continue measuring.
L
R
10
Cursor ON/OFF
Set the cursor position manually, by dragging it to the left or right. The
trigger position may also be adjusted later on the saved waveform in the
Explorer.
11
Automatic Estimation of Pulse Velocity
Ideally the pulse velocity should be entered manually after having made a
control measurement on an object of known thickness. If this is not pos-
sible, it is possible to estimate the pulse velocity directly on surface of
the test object. Tap on this button and press the transducer against the
surface to make a measurement. The surface velocity will be measured
automatically and stored as the pulse velocity for subsequent B-Scan or
distance measurements.