Specifications

7. Maintenance
7.1 General
1. Check main power cord and plug if there are abrasions or
excessive wear.
2. Check mattress cover for signs of wear or damage. Ensure mat
tress cover and tubes are stubbed together correctly.
3. Check airflow from the air hose connector. The airflow should
alternate between each connector every half-cycle time if it's in
alternating mode.
4. Check the air hoses for any kink or break. For replacement, please
contact an authorized dealer.
7.2 Fuse Replacement
1. Disconnect the plug form mains power when
a blown fuse is suspected.
2. Remove the cover of the fuse holder by
means of a mall screwdriver.
3. Insert a new fuse of the correct rating in, and
replace the cover of the fuse holder back.
The fuse should be rated as T1A/250V type.
2. Product Description
2.
1 Pump and Mattress System
1.CPR
2
.
Pump unit
3.Mattress System
4.Quick Connectors
8. Specifications
Pump Specification
Power Supply AC 120V / 60 Hz, 0.1 A
Fuse Rating T 1A, 250V
Cycle Time 8 minutes
Dimensions (LxWxH) 11.8” x 3.7” x 7.5
Weight 6.2 lbs.
Environment Temperature
Humidity
Operation: 50°F to 104°F
Storage: 5°F to 122°F
Shipping: 5°F to 158°F
Operation: 10% to 90% non-condensing
Storage: 10% to 90% non-condensing
Shipping: 10% to 90% non-condensing
Classification Class I, Type BF, IPX0
Applied Part: Air Mattress
Not suitable for use in the presence of a
flammable anesthetic mixture (No AP or
APG protection)
11 4