Operating instructions

The M&C gas converter CG-10 is designed as a compact, user-
friendly and easy to service wall mounting unit.
Gas connection for sample and purge are through the bottom of
the case as shown above.
Electrical and signal connections are through conduit entry
glands on the bottom right rear of the case as shown.
All operating elements are mounted on the front panel of the
converter for easy access. These are:
Mounting adapter with handle for the catalyst cartridge
Temperature controller with digital temperature display
Selector switch for bypassing the catalyst cartridge
The catalyst cartridge is mounted in a heat-insulated tube furnace.
The special mounting adapter with handle 2 allows the hot
catalyst cartridge to be released and removed without tools. But
the use of hot gloves for personnel protection is required if
handling a hot, freshly removed cartridge.
Use extreme caution when handling a removed cartridge to
avoid burns
The converter temperature is electronically controlled and can
be adjusted by means of the temperature controller 3 on the
converter front panel in a range between +50 C and +360 C,
Mo/C cartridges operate within a recommended temperature range of
320°C to 340°C.
Due to the non-ventilated case of the CG-10, the maximum oven set
point temperature is 360°C.
One alarm for low-temperature is provided as a status contact
output, available on the DIN rail mounted general purpose relay
socket within the unit. Form C contacts are provided.
;
M&C Produc
ts Analysis Technology 9-2.0-MEUS