Specifications

Minitrend
®
GR 10
Minimum System requirements for
TrendViewer:·
Minimum System requirements for TrendServer
Pro, TrendManager Pro and GR Series Screen
Designer:
Supports: Microsoft Windows™ Windows 7 (32 and 64 bit
– Professional, Enterprise and Ultimate Edition), Windows
2008, Windows 2008 Server, Windows 2012 Server.
Supports: Microsoft Windows™ Windows 7 (32 and 64 bit –
Professional, Enterprise and Ultimate Edition), Windows
2008, Windows 2008 Server, Windows 2012 Server.
1 GHz Pentium processor or higher with a mouse, CD-
ROM drive
1 GHz Pentium processor or higher with a mouse, CD-ROM
drive
Monitor screen resolution 1024 x 768 recommended
minimum, high colour
Monitor screen resolution 1024 x 768 recommended
minimum, high colour
512 Mbyte of RAM
4GB SRAM for generating IQOQ report
16 bit colour graphics, 24 bit recommended (Screen
Designer only)
16 bit colour graphics, 24 bit recommended (Screen
Designer only)
50 Mbyte free hard disk space 2 Gbyte Hard-drive free disk space
Flash card reader or USB port for X Series recorders Flash card reader or USB port for X Series recorders
3.5” floppy disk drive or PCMCIA for V5 recorders
3.5” floppy disk drive or PCMCIA for V5 recorders
TCP/IP Installed
Graphics Card & Direct-X control installed (Screen
Designer)
Specification Environmental and Operating Conditions
Parameter Reference Rated Extreme Transport & Storage
Ambient Temperature 67 °F to 77 °F
19 °C to 25 °C
32 °F to 122 °F
0 °C to 50 °C
32 °F to 122 °F
0 °C to 50 °C
14 °F to 140 °F
–10 °C to 60 °C
Relative Humidity (%RH) 50 to 65* 10 to 90* 5 to 90* 5 to 95*
Vibration: Frequency (Hz)
Acceleration (g)
0
0
0 to 70
0.1
0 to 100
0.2
0 to 100
0.5
Mechanical Shock
Acceleration (g)
Duration (ms)
0
0
1
30
5
30
20
30
Mounting Position (Vertical)
Tilted Forward
Tilted Backward
Tilted to Side (+/-)
40º
65º
65º
40º
65º
65º
Any
Any
Any
Power Requirements
Mains Voltage (Vrms)
Low Voltage AC (Vrms)
DC Voltages
Frequency (Hz)
120/240
18 +/- 2
24 +/- 2
49.8 to 50.2
100 to 240
12 to 20
12 to 30
47 to 63
90 to 264
12 to 20
12 to 35
47 to 63
N/A
N/A
N/A
N/A
Power Consumption AC: <40W (max), DC: <40W (max). Typical 20W
Warm Up 30 minutes minimum
Seismic Qualification Complies with IEEE 323-1974 and/or 1983 and IEEE 344-1975 and/or 1987 (optional)
* The maximum rating only applies up to 104°F (40°C). For higher temperatures, the RH spec is de-rated to maintain constant moisture