Product Manual

5
Section 4 Analyzer Setup
Press Menu from the Initial Screen to display the screen below.
Select options by pressing or until the Selection Arrow () is next to the
option. Press ENTER to show the options in a new screen or cycle through the
settings below. Select Exit and press ENTER to return to the Initial Screen.
Test Settings
Select Test Settings to display the menu shown below.
Press or to select an option. Press Enter to show or change the option
settings below. Select Exit and press Enter to return to the Main Menu.
2P, 4P, 7P, and 10SG are examples of Type of Strips. Refer to Appendix 2 for
actual Analyzer-Read Type of Strips.
Type of Strip 2P, 4P, 7P, and 10SG
Mode Routine, STAT, QC
Test Number
User selected number
Continuous Test On, Off (single test)
Operator ID
If User Login is off, Operator ID will be 01 through 10.
If User login is enabled, Operator ID will be 11
through 20, and 100 for the Administrator
Units Conventional Units or Standard International (SI)
Type of Strip
Select Type of Strip and press ENTER to change the strip type. Each strip type
name defines the number of test parameters. If User Login is enabled and
Operator ID is 11 through 20, this setting cannot be changed.
Refer to Appendix 2 for a detailed list of available parameters.
Note: Ensure the type of strip selected corresponds with the strip to be
used. If it does not, an error message will be displayed.