Section 3 Initial Startup
Place the analyzer on a level surface. Allow 80*50 cm on all sides of the analyzer
for access.
Plug the power cord into a power outlet. Press the power switch (19) located on
the back panel to turn the Pro Advantage
®
Urine Analyzer on. The Pro
Advantage
®
Urine Analyzer will run a Self-Test. The analyzer will display the
screen shown below.
“Optical Sensor Failed” will display due to lack of Strip Holder. The Strip Holder
Mount will extend for Strip Holder installation. Turn off the Power Switch.
Remove the Strip Holder from its plastic packaging. Place the Strip Holder on the
Strip Holder Mount. Position the white Calibration Circle up and towards the back
of the analyzer. It will slide loosely in the locating channel. Push the Strip Holder
in until it clicks into place. It will be held firmly on the Strip Holder Mount. Ensure
the Strip Holder is installed properly. If it is not locked onto the Strip Holder Mount
the analyzer will display “Optical Sensor Failed.”
Turn the Power Switch on. If the Self-Test passes, the Initial Screen will be
shown.
Note: The screen will not display
if Strip Lockout is not included.