The exact agreement between the CM and the WM is 86.3% (483/560) with 95% CI: (83.2%; 89%).
Agreement within +/- 1 block is 97.7% (547/560) with 95% CI: (96.1%; 98.8%).
Leukocytes
Bayer Clinitek 500 Urine Chemistry Analyzer
0 15 70 125 500 Total
0
268 23 4 0 0 295
15
14 34 17 0 0 65
70
8 10 53 7 3 81
125
1 0 4 80 22 107
Pro
Advantage
®
Urine
Analyzer
500
0 0 0 1 11 12
Total
291 67 78 88 36 560
The exact agreement between the CM and the WM is 79.6% (446/560) with 95% CI: (76.0%;
82.9%). Agreement within +/- 1 block is 90.5% (507/560) with 95% CI: (87.8%; 92.8%).
Protein
Bayer Clinitek 500 Urine Chemistry Analyzer
0 15 30 100 300 Total
0
265 10 0 0 0 275
15
19 42 13 0 0 74
30
0 9 48 17 0 74
100
0 0 0 42 9 51
Pro
Advantage
®
Urine
Analyzer
300
0 0 0 0 86 86
Total
284 61 61 59 95 560
The exact agreement between the CM and the WM is 86.3% (483/560) with 95% CI: (83.2%; 89%).
Agreement within +/- 1 block is 94.8% (531/560) with 95% CI: (92.7%; 96.5%).
Ketone
Bayer Clinitek 500 Urine Chemistry Analyzer
0 5 15 40 80 Total
0
317 10 0 0 0 327
5
5 55 9 0 0 69
15
0 0 52 21 0 73
40
0 0 6 46 2 54
Pro Advantage
®
Urine
Analyzer
80
0 0 0 9 28 37
Total
322 65 67 76 30 560
The exact agreement between the CM and the WM is 88.9% (498/560) with 95% CI: (86.0%;
91.4%). Agreement within +/- 1 block is 97.3% (545/560) from each other with 95% CI: (95.6%;
98.5%).