Product Manual

35
Section 10 Troubleshooting
Problem Solutions
Strip Position Improper
Adjust the strip so the strip end is completely touching the Strip Holder
Backstop and centered in the Strip Holder Channel
Strip Missing Insert strip
Strip Error
Ensure the type of strip used matches the type of strip setting
Ensure the strip brand is compatible with the analyzer
Ensure all of the test pads on the strip have been immersed in the
specimen
No display on screen
Turn the Power Switch on
Make sure power is applied to the Analyzer
Examine the Fuse on the back of the analyzer to determine if it is
damaged and replace if necessary
Fuse is damaged
Turn the Power Switch off, disconnect the plug and replace the damaged
Fuse with a new 2.0A Fuse
(Spare fuse is located in the fuse compartment)
Printer does not work
Make sure Auto-Print is set to On. Refer to Printer Setup instructions
Load paper if necessary. Ensure Printer Setup is set for Internal when an
external printer is not connected
Ensure paper roll is placed correctly. If incorrect, there will be no printout
Clean printer roller with ethanol
See that the proper side of the thermal paper is inserted in the printer
Remove paper jam by lifting up green paper release lever and removing
jammed paper. (See Section 2)
Main Control Unit Failed
Turn the Power Switch off, then on. Perform a Self-Test. Refer to
Manual Self-Test section
Optical Sensor Failed
Ensure the Strip Holder is locked onto the Strip Holder Mount correctly.
The Strip Holder should snap into place and locked onto the Strip Holder
Mount.
Ensure the white Calibration Circle is clean
Clean the Strip Holder and refer to Daily Cleaning section
Turn the Power Switch off, then on. Perform a Self-Test. Refer to
Manual Self-Test section
Test Accuracy Failed
Turn the Power Switch off, then on. Perform a Self-Test. Refer to
Manual Self-Test section
Mechanism Failed
Remove any obstacles in the path of the Strip Holder
Do not touch the Strip Holder when it is moving
Turn the Power Switch off, then on. Perform a Self-Test. Refer to
Manual Self-Test section
Excess Light Failed
Ensure the analyzer is kept away from direct light
Ensure the white calibration circle is clean
Ensure the Strip Holder is positioned correctly on the Strip Holder Mount.
Refer to Daily Cleaning section
Turn the Power Switch off, then on. Perform a Self-Test. Refer to
Manual Self-Test section
Canister Code Error
Ensure the canister code entered is correct
Ensure the type of strip used is the same as the analyzer setting.
QC Test Fail
Ensure the control is correct
Ensure the type of strip is correct
Ensure the brand of strip is compatible with the analyzer
Ensure all of the reagent pads of the strip have been immersed
For US customers, call customer service toll free at 1-(800)-838-9502 for details. For customers
outside the US, contact your local distributor.