Product Manual

30
Section 7 Quality Control
Each lab should use its own standard and procedures for performance. Test
known positive and negative specimens/controls at the following events in
accordance with local, state, and/or federal regulations or accreditation
requirements.
A new canister of strips is opened
A new operator uses the analyzer
Test results seem inaccurate
After performing maintenance or service on the analyzer
If the QC tests do not provide expected results, perform the following checks:
Ensure the strips used are not past their expiration date.
Ensure strips are fresh from a new canister.
Ensure the controls are not past their expiration date.
Repeat the test to ensure no errors were made during the test.