Product Manual
27
QC lockout
When QC lockout is turned On, the user will be notified when a QC test is required.
The following screen will be displayed. The QC symbol will flash.
Note: Use Pro Advantage
®
Urine Reagent Strips for proper functioning
and accurate results.
If it is the first time to run the QC test, refer to the section on
Setting the QC Values to set up QC test values.
Press CANCEL to return to the Initial Screen. If canceled, the analyzer will
change the test mode to STAT. An “E” will be displayed after the test number on
all printouts to show QC tests were out of date and pending.
Press ENTER to run QC tests.
The analyzer will run a Level 1 QC test first. Refer to the Strip Test section for
details on testing.
After testing, the analyzer will display the testing results.
If any parameter is out of range, it will be highlighted and marked with a “*”.
If Auto-print is set to On, the result will be printed.
Press ENTER to test control Level 2. The steps are the same as for Level 1.