Product Manual
26
When the test is complete, the Results Screen will show briefly. The next strip
can be placed on the Strip Holder.
Stop testing by pressing CANCEL when the prompt appears. All other testing
procedures are the same as testing with no barcode.
Urine Controls QC Testing
Ensure the operating Mode is set to QC. All test numbers in QC mode will begin
with 2. This allows results to be searched for and found easily.
Strip Preparation
Allow the strip and urine controls to reach room temperature at 15-30ºC (59-86ºF)
prior to testing.
Note: Use Pro Advantage
®
Urine Reagent Strips for proper functioning
and accurate results.
Remove strips from the closed canister and use them as soon as possible. Close
the canister tightly immediately after removing the strips.
Urine Control Test Procedures
The urine control test procedures are the same as Normal Operation, no Barcode
reader, or Normal Operation, Barcode reader installed. Refer to Section 5
Analyzer Operation.
The results obtained during the Quality Control test will be referenced by the
analyzer with pre -programmed target values and report generated, If Auto-print is
set to On, the result will be printed.
If the QC test "Fails", please contact your local distributor for Technical Support.