Product Manual
22
Note: Timing is set up to process a new strip every 30 seconds.
Once START is pressed, the screen will change to the following:
The countdown clock is displayed on the bottom right of the screen. The clock will
start to count down from 65. The operator has 3 seconds before the triple beep
sounds to immerse the strip into the urine.
Upon hearing the triple beep and/or seeing the countdown clock reach 62,
completely immerse the reagent areas of the strip in fresh, well-mixed urine.
Immediately remove the strip to avoid dissolving the reagents. Handle and test
the strips as described in Strip Tests - Single Test Mode above.
Place the first strip onto the Strip Holder Channel. Ensure the reagent pads face
up. Make sure the end of the strip touches the Strip Holder Backstop.
Listen for the next triple beep and/or see the countdown clock reach 32. Immerse
a new second strip in fresh, well-mixed urine. Repeat the strip testing steps
above. Lay the second urine-exposed strip on a paper towel, with reagent pads
up. Wait for the first strip to complete testing and the Strip Holder to move out.
Listen for the next triple beep and/or see the countdown clock reach 2. Immerse
a new third strip in fresh, well-mixed urine. Repeat the strip testing steps above.
Lay the third strip on a paper towel with reagent pads up. Wait for the Strip Holder
to move out.
Note: When the countdown clock reaches 1, the Strip Holder will carry the
first strip inside. The countdown clock also returns to 30. The
results will be displayed on the screen for several seconds and
stored in memory. If Auto-print is on, the results will be printed.
Remove the completed first strip after the Strip Holder moves out. Immediately
place the second strip on the Strip Holder before the countdown clock reaches 1.
The Strip Holder will carry the new strip inside for analysis.
Note: The operator has about 8 seconds to discard the used strip and
place the next strip on the Strip Holder.