Product Manual

21
Run the edge of the strip against the rim of the urine specimen container to remove
excess urine. Hold the strip in a horizontal position. Bring the edge of the strip
into contact with an absorbent material (e.g. a paper towel). This prevents mixing
chemicals from adjacent reagent areas.
Place the strip with the reagent area facing up, onto the Strip Holder Channel.
Make sure the strip end touches the Strip Holder Backstop as shown below.
Note: When the countdown clock on the display reaches 1, the Strip
Holder will carry the strip inside and begin testing.
An automatic calibration is done before each test.
The results will be displayed on the screen and stored in memory
after each test. Any abnormal results will be highlighted on the
screen and flagged on the print out.
If Auto-print is set to on the results will be printed. If Auto-print
is set to off, press Print to print the results.
Remove the used strip from the Strip Holder when the strip carrier moves out.
Discard the used strip according to local regulations.
Repeat the previous steps to test more urine.
Strip Test – Continuous Test Mode
Press START when the Strip Holder is fully extended to begin the test. The
following screen will be displayed. Wait for the audible triple beep to immerse the strip.