Product Manual

20
that a new canister code is required, scan or manually enter a
canister code from a new canister to be used, or from the one
currently in use. The analyzer will not accept canister codes that
have already been entered and will lockout the user until a new
canister code is entered. One canister code can be entered for
each Type of Strip, and will be managed independently of other
strip types by the analyzer.
Normal Operation, no Barcode Reader
Sample/Strip Preparation
Allow the strip, urine specimen, and/or controls to reach room temperature at
15-30ºC (59-86ºF) prior to testing.
Note: Only Pro Advantage
®
Urine Reagent Strips work with the Pro
Advantage
®
Urine Analyzer. Use of any other strips will cause
inaccurate results.
Remove the strips from the closed canister. Use them as soon as possible.
Tightly close the canister after removing the strips.
Strip Test – Single Test Mode
Press ENTER for the strip prompt. Wait for the audible triple beep to immerse the
strip.
Once START is pressed, the screen below will show.
The countdown clock is displayed on the bottom right. The clock will start to count
down from 65. The operator has 3 seconds before the triple beep sounds to
immerse the strip into the urine.
Upon hearing the triple beep and/or seeing the countdown clock reach 62,
completely immerse the reagent areas of the strip in fresh, well-mixed urine.
Immediately remove the strip to avoid dissolving the reagents.