Product Manual
19
At first startup, the Strip Holder will fully extend. The screen below will be
displayed.
Scan or manually enter the canister code from the strip canister. Press CLEAR to
move the cursor to the proper digit. Edit the number as needed. After the last
digit of the correct code is entered, the next screen will be shown. Press
CANCEL to cancel the changes and exit the screen.
If the canister code is incorrect, the following screen will show.
Press CANCEL to exit to the previous screen.
If the canister code is correct, the analyzer will proceed to the Initial Screen.
Note: Refer to Section 4, Analyzer Setup, to change the settings.
When there are only 5 strips remaining,
will flash to prompt that a new canister
of strips is required. The number of strips left can be seen under
.
When the current run is completed, enter the new canister code. The number
under
will be increased accordingly.
Note: Refer to the following for more information on the Strip
Lockout function. Once a canister code is accepted, the
analyzer will automatically recognize the Type of Strip and the
number of strips per canister should be tested. The analyzer will
keep track of the number of strips available for testing and the
strips remaining will be displayed under the lock symbol.
Continue testing as usual until the analyzer prompts that a
new canister code is required. When the analyzer prompts