Product Manual

13
QC Lockout
Select QC Lockout and press ENTER to cycle the options Off or On.
Off disables the QC lockout function. The QC symbol and Setup Options are not
displayed.
On enables the QC lockout function. The QC symbol is displayed. QC Lockout
timing can be set for Every 8 Hours, Daily, Weekly or Monthly.
Time
Select Time and press ENTER to set the QC test for Every 8 Hours, Daily,
Weekly or Monthly.
Time Setup
Select Time Setup and press ENTER.
Every 8 Hours / Daily Weekly / Monthly
The first highlighted digit will be changed. Press or to change. Continue to
press or until the correct number is shown. Press or to progress to the
next digit to be changed. Press or to increase or decrease the number.
Change each of the time and date digits/fields in this manner until the desired date
is shown. Press ENTER to save the changes and exit the screen. Press
CANCEL to not accept changes and exit.
There is no Date Setup if QC tests are set for Every 8 Hours or Daily.
For the Monthly option, Date can be set from 01 to 28. 29, 30
and 31 are invalid.
Run QC Test
Select Run QC Test and press ENTER to perform a QC test before the next
scheduled test time.
When a QC test is run for the first time, the analyzer will show the screen below if
QC values have not been set up.