Product Manual

10
Set Date and Time
Select Set Date and Time and press ENTER to display a new screen for setting
the current date and time.
The first digit to be changed in the date will be highlighted. Press or to
change the number one at a time. Continue to press or until the correct
number is shown. Press or to progress to the next digit to be changed.
Press or to increase or decrease the number. Change each of the time and
date digits/fields in this manner. Press ENTER to save the changes and exit the
screen. Press CANCEL to not accept any changes and exit the screen. Either
key will return to the previous screen.
Auto Number Reset
Select Auto Number Reset and press Enter to cycle to Yes or No. If Yes, the
test number will reset to 000001, 100001 or 200001 for Routine, STAT or QC
modes each day. If No the test number is unaffected by date.
Database
Select Database and press ENTER to display the screen below.
Press or to move to enter each screen and make changes. After all
changes are made press ENTER to save changes and return to the Main Menu.
Up to 2000 records can be stored in memory. If User Login is enabled and
Operator ID is 11 through 20, the operator can only review data stored in the
analyzer.
Caution: Use a number higher than the previous test number.
If there are 2000 records stored in memory, the oldest record will be erased and
replaced.
Records may have the same number but different dates and times.