Quick Guide
iPRECIO
®
IMS-310R Quick guide ver.1.0.0.E
24
www.iprecio.com
environment.
Verify PC and UCD-X10R connect properly. For detail, please refer “TCP/IP
connection between data com unit and control PC” and “Network Setting”.
Verify pump activation. For detail, please refer “Catheter connection,
plomb admin liquid and activation”.
3. After all pumps detected, then click Next button.
KVO setting, start and complete
KVO is abbreviate of Keep Vein Open, semantics to prevent the catheter edge clog by
continuous liquid dosing by low flow rate. Be normally administrate biocompatible
liquid such as saline as KVO by low flow rate after implant surgery and prevent clog
of catheter end during animal healing period. After animal comeback form the surgery,
the KVO liquid in reservoir can be replaced by medicine liquid from the septum. And
acknowledge user the timing that when the medicine liquid will reach the catheter end
by pump and appurtenant catheter dead volume calculation.
1. Open desire Group profile and check the KVO check BOX and click the dead volume
setting button. Then KVO setting window pops up. If KVO is not required, then
remove check the KVO check box to skip KVO.
2. Click individual column dead volume setting.
3. Input the required information for dead volume calculation and click OK button.