Quick Guide
iPRECIO
®
IMS-310R Quick guide ver.1.0.0.E
28
www.iprecio.com
Replace and/or Refill the medical solution
1. Open Monitor (M) window.
2. Verify the estimated remain volume in the bottom area.
3. Make pump implanted animal immobilization by weak anesthesia or retention.
4. When replace the solution, aspirate remain fluid in reservoir by 27G syringe needle
through the skin to port. Subsequently fill the replacing solution into syringe and
refill it to reservoir. To restock the drag solution, just fill the solution in a loaded
syringe through the skin to port.
5. Input Refill Time and Volume, click Enter button. When executed by dose mode,
input new weight and click Calculate button, recast New Concentration result will
be displayed to keep dose condition.
6. After fix the Refill Log, the table record automatically the Refill time, volume and
comments (weight and density). To remove the refill log, select target refill log and
click “Delete” button.