Even if there is no need to change dead volume setting, it need to open every
animal dead volume setting and determine the dead volume choice. If this
determine was lost, the dead volume and flash rate will not apply to the
administration start schedule.
4. Check KVO flow rate and KVO duration and if it’s ok then click to KVO start. The
pump status will be KVO after start KVO. Pump status color will change in
accordance with communication stats, red Yellow Green. When the color will
be Green, then injection is started.
During KVO is running, please DO NOT STOP the management software.
5. When the pre-set KVO period will end, pumps will stop and Pump status changes
into Finished. Click Next>, the injection schedule setting. All of animal looks fully
recovered before pre-setting KVO period, you can abort all KVO by clicking abort
as a pack and start administration schedule. When some animal cannot heal
enough during KVO period and need to terminate the experiment of that animal,
then click the remove button from individual animal column.
After KVO stopped, please check the Pump status as “Finish”. Demepnd on the
communication condition, “Finish” status appears very slow after prompt to
end KVO from management software due to communication condition.