Quick Guide
iPRECIO
®
IMS-310R Quick guide ver.1.0.0.E
22
www.iprecio.com
Catheter connection, plomb admin liquid and activation.
Activate the pump by switch on to communication enable. If required connect
additional catheter as necessary.
To fill administration liquid till the end of catheter before power on, because
once a pump switched on, the internal cam will start moving to the initial
position to block the inner tube.
1. Open pump pre-sterilization pack in clean environment.
2. Cut, replace tube and/or catheter connection of pump as necessary.
Each pump dead volume can be calculated in KVO setting, so keep final tube
length and inner diameter or measured dead volume need to be recorded.
3. Fill the admin into the end of tube with 27G syringe needle through upper port.