HELIO2
BLENDEr
HELIUM-OXYGEN
WARNING
Only trained, qualied medical personnel under the direct supervision of a
licensed physician should operate the
HELIO2
Blender.
Use this HELIO2 Blender only for its Intended Use as described in this
manual.
Conrm prescribed dose before administering to patient. Monitor on a
frequent basis. Continuous Monitoring with an alarmed Oxygen Monitor
/ Analyzer is recommended.
The
HELIO2
Blender shall be serviced by a qualied service technician.
Always follow ANSI and CGA standards for Medical Gas Products,
Flowmeters and Oxygen Handling.
The clinician must verify the Heliox tank concentration prior to utilizing
the Blender.
An alarmed Oxygen Monitor / Analyzer must be used to verify oxygen
concentration.
Accuracy of oxygen concentration will be affected if bleed is not activated
at ow settings below 15 l/min for the High Flow Blender, and 3 l/min for
the Low Flow Blender.
• DO NOT obstruct the alarm.
• DO NOT use Blender when alarm is sounding.
• DO NOT use oil in or around the Blender.
• DO NOT occlude or obstruct the bleed port on the auxiliary outlet of
the Blender.
• DO NOT use near any type of ame or ammable/explosive
substances, vapors or atmosphere.
• Oxygen Concentration Dial does not rotate 360 degrees.
Rotating the
dial beyond the endpoint settings will damage the Blender.
CAUTION
• Turn off gas supplies when
HELIO2
Blender is not in use.
• Store the
HELIO2
Blender in a clean, dry area when not in use.
• The HELIO2 Blender contains magnetic, ferrous material that may
affect the results of an MRI.
• Ensure all connections are tight and leak free.
• Avoid excessive pressure surges greater than 100 psi (6.9 bar) when
pressuring the Blender inlets.
• DO NOT steam autoclave.
• DO NOT immerse
HELIO2
Blender into any liquid.
• DO NOT gas sterilize with (EtO) Ethylene Oxide.
• DO NOT use if dirt or contaminants are present on or around the
Blender or connecting devices.
• DO NOT smoke in an area where oxygen is being administered.
• DO NOT clean with aromatic hydrocarbons.
• Inlet pressure of Device used in conjunction with the
HELIO2
Blender must
match inlet pressure of
FIO2
Blender.
• When using a bottled high pressure gas source, always use a pressure
reducing regulator set within 30-75 psi (2.1-5.2 bar).