User Manual
5
Reassembly
1. Remove 4 plugs from new sieve bed assembly and slide it into the concentrator.
2. Gently push until sieve bed replacement is fully seated, flush with bottom surface. Replace bottom cover
and secure screws.
3. Turn unit upright and insert battery. Gently push until battery is locked in place.
Registration
1. Press and hold the Audio Pause button to enter the menu.
2. Press the 5 button to navigate to the Sieve Beds option.
3. Press the 3 button to select the Sieve Beds option.
4. Press the 3 button again to register the sieve bed assembly.
5. “Done” will appear on the screen once completed.
6. Press the Audio Pause button to exit the menu.